For a CHAMBER ASSEMBLY with an associated MEASURING ASSEMBLY it is the factor which converts the INDICATEDVALUE, corrected to stated REFERENCE CONDITIONS, to the CONVENTIONAL TRUE VALUE at the position of theREFERENCE POINT of the IONIZATION CHAMBER. For an IONIZATION CHAMBER calibrated on its own without a specifiedMEASURING ASSEMBLY, the CALIBRATION FACTOR converts the ionization charge or current, corrected to REFERENCECONDITIONS, to the CONVENTIONAL TRUE VALUE at the position of the REFERENCE POINT of the CHAMBER (this is thereciprocal of RESPONSE under REFERENCE CONDITIONS).
Za SKLOP JONIZACIONE KOMORE sa pripadajućim MERNIM SKLOPOM, faktor koji PRIKAZANU VREDNOST, korigovanu prema postojećim REFERENTNIM USLOVIMA, pretvara u KONVENCIONALNU PRAVU VREDNOST na mestu REFERENTNE TAČKE JONIZACIONE KOMORE. Za JONIZACIONU KOMORU koja je kalibrisana sama bez specificiranog MERNOG SKLOPA, KOEFICIJENT ETALONIRANJA pretvara jonizaciono naelektrisanje ili struju, korigovane prema REFERENTNIM USLOVIMA, u KONVENCIONALNU PRAVU VREDNOST na mestu REFERENTNE TAČKE KOMORE (to je recipročna vrednost ODZIVA U REFERENTNIM USLOVIMA).
Za SKLOP JONIZACIONE KOMORE sa pripadajućim MERNIM SKLOPOM, faktor koji PRIKAZANU VREDNOST, korigovanu prema postojećim REFERENTNIM USLOVIMA, pretvara u KONVENCIONALNU PRAVU VREDNOST na mestu REFERENTNE TAČKE JONIZACIONE KOMORE. Za JONIZACIONU KOMORU koja je kalibrisana sama bez specificiranog MERNOG SKLOPA, KOEFICIJENT ETALONIRANJA pretvara jonizaciono naelektrisanje ili struju, korigovane prema REFERENTNIM USLOVIMA, u KONVENCIONALNU PRAVU VREDNOST na mestu REFERENTNE TAČKE KOMORE (to je recipročna vrednost ODZIVA U REFERENTNIM USLOVIMA).