ISO/FDIS 10993-7

Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals

Опште информације

50.00     2. 12. 2024.

ISO

ISO/TC 194

Међународни стандард

11.100.20  

Апстракт

ISO 10993-7:2008 specifies allowable limits for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) in individual EO-sterilized medical devices, procedures for the measurement of EO and ECH, and methods for determining compliance so that devices may be released. Additional background, including guidance and a flowchart showing how the standard is applied are also included in informative annexes.
EO-sterilized devices that have no patient contact (e.g., in vitro diagnostic devices) are not covered by ISO 10993-7:2008.

Животни циклус

ПРЕТХОДНО

ОБЈАВЉЕН
ISO 10993-7:2008

ОБЈАВЉЕН
ISO 10993-7:2008/Cor 1:2009

ОБЈАВЉЕН
ISO 10993-7:2008/Amd 1:2019

ТРЕНУТНО

ПРОЈЕКАТ
ISO/FDIS 10993-7
50.00 Евидентирање података о дефинитивном тексту нацрта стандарда
2. 12. 2024.

Национална преузимања

Биолошко вредновање медицинских средстава – Део 7: Остаци после стерилизације етилен-оксидом

45.99   Слање нацрта FV у CCMC