Phone: (011) 7541-421, 3409-301, 3409-335, 6547-293, 3409-310
E-mail: Standards sales: prodaja@iss.rs Education: iss-edukacija@iss.rs Information about standards: infocentar@iss.rs
Stevana Brakusa 2, 11030 Beograd
Main menu

SRPS EN ISO 19211:2025

Anaesthetic and respiratory equipment - Fire-activated oxygen shut-off devices for use during oxygen therapy (ISO 19211:2024)

Oct 31, 2025

General information

60.60     Oct 31, 2025

ISS

Z076

European Norm

11.040.10  

English  

Buying

Published

Language in which you want to receive the document.

Scope

This document specifies requirements for fire activated oxygen shut-off devices that stop the flow of oxygen in respiratory therapy tubing when activated by fire.
NOTE 1        Typical arrangements for fire activated oxygen shut-off devices are shown in Annex C.
NOTE 2        Respiratory therapy tubing is covered by ISO 17256.
NOTE 3        Use of fire activated oxygen shut-off devices in medical devices or accessories is not mandated in this document.
The fire activated oxygen shut-off devices specified in this document are not suitable for use with oxygen therapy systems with flows in excess of 20 l/min).
NOTE 4        There is rationale for this clause in A.2.
The requirements in this device-specific standard take precedence over any conflicting requirements in the general standard for airway devices (ISO 18190). All the common requirements that appear in the general standard for airway devices have been removed from this document.

Life cycle

NOW

PUBLISHED
SRPS EN ISO 19211:2025
60.60 Standard published
Oct 31, 2025

Related project

Adopted from EN ISO 19211:2025 IDENTICAL

Adopted from ISO 19211:2024 IDENTICAL