prSRPS EN ISO 14801:2025

Dentistry — Implants — Dynamic loading test for endosseous dental implants

General information

10.99     Sep 11, 2025

ISS

Z106

European Norm

English  

Scope

ISO 14801:2016 specifies a method of dynamic testing of single post endosseous dental implants of the transmucosal type in combination with their premanufactured prosthetic components. It is most useful for comparing endosseous dental implants of different designs or sizes. This International Standard is not a test of the fundamental fatigue properties of the materials from which the endosseous implants and prosthetic components are made.
This International Standard is not applicable to dental implants with endosseous lengths shorter than 8 mm nor to magnetic attachments.
While ISO 14801:2016 simulates the functional loading of an endosseous dental implant under "worst case" conditions, it is not applicable for predicting the in vivo performance of an endosseous dental implant or dental prosthesis, particularly if multiple endosseous dental implants are used for a dental prosthesis.

Life cycle

PREVIOUSLY

PUBLISHED
SRPS EN ISO 14801:2017

NOW

PROJECT
prSRPS EN ISO 14801:2025
10.99 New project approved
Sep 11, 2025

Related project

Adopted from prEN ISO 14801 rev IDENTICAL

Adopted from ISO/AWI 14801 IDENTICAL