Published
ISO 11140-3:2007 specifies the requirements for chemical indicators to be used in the steam penetration test for steam sterilizers for wrapped goods, e.g. instruments and porous materials. The indicator for this purpose is a Class 2 indicator as described in ISO 11140-1.
Indicators complying with ISO 11140-3:2007 are intended for use in combination with the standard test pack as described in EN 285. ISO 11140-3:2007 does not consider the performance of the standard test pack, but does specify the performance of the indicators.
NOTE: If the word "harmonized" (marked in green) is not found in the field with the name of the directive, it means that the European standard is not cited in the OJEU.
Directive 2007/47/EC of the European Parliament and of the council of 5 September 2007 amending Council Directive 90/385/EEC on the approximation of the laws of the Member States relating to active implantable medical devices, Council Directive 93/42/EEC concerning medical devices and Directive 98/8/EC concerning the placing of biocidal products on the market
WITHDRAWN
SRPS EN ISO 11140-1:2009
PUBLISHED
SRPS EN ISO 11140-3:2011
60.60
Standard published
May 31, 2011