Published
ISO 11137-2:2013 specifies methods for determining the minimum dose needed to achieve a specified requirement for sterility and methods to substantiate the use of 25 kGy or 15 kGy as the sterilization dose to achieve a sterility assurance level, SAL, of 10−6. ISO 11137-2:2013 also specifies methods of sterilization dose audit used to demonstrate the continued effectiveness of the sterilization dose.
ISO 11137-2:2013 defines product families for sterilization dose establishment and sterilization dose audit.
WITHDRAWN
ISO 11137-2:2012
PUBLISHED
ISO 11137-2:2013
90.92
Standard to be revised
Jul 23, 2024
PUBLISHED
ISO 11137-2:2013/Amd 1:2022
PROJECT
ISO/CD 11137-2
Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose (ISO 11137-2:2013)
95.99 Withdrawal of Standard
Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose (ISO 11137-2:2013)
60.60 Standard published