Withdrawn
ISO 10993-3:2003 specifies strategies for hazard identification and tests on medical devices for the following biological aspects: genotoxicity, carcinogenicity, and reproductive and developmental toxicity.
ISO 10993-3:2003 is applicable for evaluation of a medical device whose potential for genotoxicity, carcinogenicity or reproductive toxicity has been identified. 
Guidance on selection of tests is provided in ISO 10993-1.
NOTE: If the word "harmonized" (marked in green) is not found in the field with the name of the directive, it means that the European standard is not cited in the OJEU.
                    WITHDRAWN
                    SRPS EN ISO 10993-3:2008
                    95.99
                                        Withdrawal of Standard
                    Sep 30, 2011
                
                WITHDRAWN
                                    SRPS EN ISO 10993-3:2011