SRPS EN 13612:2008

Performance evaluation of in vitro diagnostic medical devices

Oct 9, 2008

General information

90.60     Sep 2, 2025

ISS

Z076

European Norm

11.100  

English  

Buying

Published

Language in which you want to receive the document.

Scope

This European Standard applies to the performance evaluation of in vitro diagnostic medical devices (IVD MDs) including IVD MDs for self-testing. It specifies the responsibilities and general requirements for the planning, conduct, assessment and documentation of a performance evaluation study by the manufacturer. It does not apply to specific evaluation plans for certain IVD MDs or a specific use.

NOTE For a selection of publications on specific evaluation plans see Bibliography.

Where a manufacturer maintains a quality system this standard addresses the compliance with "design validation" and "design changes" as described in EN ISO 9001, EN 46001 and EN 928 especially considering the nature and use of IVD MDs.

In particular, this standard applies to IVD MDs to

- show evidence to notified bodies and national authorities by results of a performance evaluation that the IVD MD performs as claimed by the manufacturer,

- establish adequate performance evaluation data originating from appropriate studies or resulting from available literature, and to

- satisfy the requirements of a quality system for design validation.

Related directives

NOTE: If the word "harmonized" (marked in green) is not found in the field with the name of the directive, it means that the European standard is not cited in the OJEU.

98/79/EC

In vitro diagnostic medical devices

Harmonized

Life cycle

NOW

PUBLISHED
SRPS EN 13612:2008
90.60 Close of review
Sep 2, 2025

Related project

Adopted from EN 13612:2002