Published
Specifies two methods for the determination of cadmium: using atomic absorption spectrometry (AAS) in an air-acetylene flame (aspiration of the acidified sample into the flame and measurement of the cadmium concentration at a wavelength of 228,8 nm) and by electrothermal atomization AAS (injection of the acidified sample into an electrically heated graphite tube of an electrothermal atomization atomic absorption spectrometer and measurement of the absorbance at a wavelength of 228,8 nm).
NOTE: If the word "harmonized" (marked in green) is not found in the field with the name of the directive, it means that the European standard is not cited in the OJEU.
Pollution caused by certain dangerous substances discharged into the aquatic environment of the Community
WITHDRAWN
SRPS ISO 5961:1997
PUBLISHED
SRPS EN ISO 5961:2009
90.93
Standard confirmed
May 21, 2008