SRPS EN 61010-2-081:2015

Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes

Jul 28, 2015
95.99   Withdrawal of Standard   Nov 30, 2023

General information

95.99     Nov 30, 2023

ISS

N045

European Norm

71.040.10     19.080  

English  

Buying

Withdrawn

Language in which you want to receive the document.

Scope

<span style='color:red'>NEW!</span> IEC 61010-2-081:2015 is available as <a href="http://webstore.iec.ch/webstore/webstore.nsf/Artnum_PK/9999966559?opendocument">IEC Standards+ 61010-2-081:2015 which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.


IEC 61010-2-081:20 15 applies to automatic and semi-automatic laboratory equipment for analysis and other purposes. Automatic and semi-automatic laboratory equipment consists of instruments or systems for measuring or modifying one or more characteristics or parameters of samples, performing the complete process or parts of the process without manual intervention. Equipment forming part of such a system is within the scope of this standard. This second edition cancels and replaces the first edition published in 2003. It constitutes a technical revision and includes the following signif icant changes from the first edition, as well as numerous other changes:

- excluded IEC 61010-2-101 (IVD Equipment) from scope. This separates IEC 61010-2-081 and IEC 61010-2-101 equipment;

- added biological risks symbol to Table 1 in Clause 5;

- added requirement for statement on hazardous substances and gases in Instructions for Use to Clause 5;

- Added marking requirement for flow direction to Clause 5;

- added requirement for OPERATOR maintenance instructions to Clause 7;

- excluded equipment whose size and weight make unintentional movement unlikely from drop test in Clause 8;

- added requirement for biohazard marking to Clause 13;

- added requirement for interlock systems containing electric/electronic or programmable components to Clause 15.

It has the status of a group safety publication in accordance with IEC Guide 104.

Related directives

NOTE: If the word "harmonized" (marked in green) is not found in the field with the name of the directive, it means that the European standard is not cited in the OJEU.

2006/95/EC

Directive 2006/95/EC of the European Parliament and of the Council of 12 December 2006 on the harmonisation of the laws of Member States relating to electrical equipment designed for use within certain voltage limits (codified version)

Harmonized
2014/35/EU

Directive 2014/35/EU of the European Parliament and of the Council of 26 February 2014 on the harmonisation of the laws of the Member States relating to the making available on the market of electrical equipment designed for use within certain voltage limits

Harmonized

Life cycle

NOW

WITHDRAWN
SRPS EN 61010-2-081:2015
95.99 Withdrawal of Standard
Nov 30, 2023

Related project

Adopted from EN 61010-2-081:2015