SRPS EN ISO 13163:2019

Water quality - Lead-210 - Test method using liquid scintillation counting (ISO 13163:2013)

Sep 24, 2019
95.99   Withdrawal of Standard   Jun 30, 2022

General information

95.99     Jun 30, 2022

ISS

H147

European Norm

13.060.60     17.240  

English  

Buying

Withdrawn

Language in which you want to receive the document.

Scope

ISO 13163 specifies the determination of lead-210 (210Pb) activity concentration in samples of all types of water using liquid scintillation counting (LSC). For raw and drinking water, the sample should be degassed in order to minimize the ingrowth of 210Pb from radon-222 (222Rn).
Using currently available liquid scintillation counters, this test method can measure the 210Pb activity concentrations in the range of less than 20 mBq⋅l-1 to 50 mBq⋅l-1. These values can be achieved with a counting time between 180 min and 720 min for a sample volume from 0,5 l to 1,5 l.
Higher 210Pb activity concentrations can be measured by either diluting the sample or using smaller sample aliquots or both.
It is the laboratory's responsibility to ensure the suitability of this test method for the water samples tested.

Life cycle

NOW

WITHDRAWN
SRPS EN ISO 13163:2019
95.99 Withdrawal of Standard
Jun 30, 2022

REVISED BY

PUBLISHED
SRPS EN ISO 13163:2022

Related project

Adopted from EN ISO 13163:2019

Adopted from ISO 13163:2013