TYPE B APPLIED PART

English

Name
TYPE B APPLIED PART
Description

APPLIED PART complying with the specified requirements of this standard IEC 60601-1/A2:1995 to provideprotection against electric shock, particularly regarding allowable LEAKAGE CURRENT and marked with symbol 1,Table DII, of Appendix D.NOTE TYPE B APPLIED PARTS are not suitable for DIRECT CARDIAC APPLICATION.

Serbian

Name
APLIKACIONI DEO TIPA B (TYPE B APPLIED PART)
Description

APLIKACIONI DEO koji je usaglašen sa specificiranim zahtevima standarda IEC 60601-1/A2:1995 za obezbeđivanje zaštite od električnog udara, posebno u odnosu na dozvoljenu STRUJU CURENJA, a koji je označen simbolom 1 u tabeli DII Priloga D.
NAPOMENA Aplikacioni delovi tipa B nisu podesni za direktnu kardiološku primenu.

Serbian

Name
APLIKACIONI DEO TIPA B (TYPE B APPLIED PART)
Description

APLIKACIONI DEO koji je usaglašen sa specificiranim zahtevima standarda IEC 60601-1/A2:1995 za obezbeđivanje zaštite od električnog udara, posebno u odnosu na dozvoljenu STRUJU CURENJA, a koji je označen simbolom 1 u tabeli DII Priloga D.
NAPOMENA Aplikacioni delovi tipa B nisu podesni za direktnu kardiološku primenu.