Telefon: (011) 7541-421, 3409-301, 3409-335, 6547-293, 3409-310
E-mail: Prodaja standarda: prodaja@iss.rs Seminari, obuke: iss-edukacija@iss.rs Informacije o standardima: infocentar@iss.rs
Stevana Brakusa 2, 11030 Beograd
Glavni meni

ISO 16063-11:1999

Methods for the calibration of vibration and shock transducers — Part 11: Primary vibration calibration by laser interferometry
16. 12. 1999.

Опште информације

90.20     15. 1. 2026.

ISO

ISO/TC 108

Međunarodni standard

17.160  

engleski  

Kupovina

Objavljen

Jezik na kome želite da primite dokument.

Apstrakt

This part of ISO 16063 specifies the instrumentation and procedure to be used for primary vibration calibration of
rectilinear accelerometers (with or without amplifier) to obtain magnitude and phase lag of the complex sensitivity
by steady-state sinusoidal vibration and laser interferometry.
It is applicable to a frequency range from 1 Hz to 10 kHz and a dynamic range (amplitude) from 0,1 m/s2 to
1 000 m/s2 (frequency-dependent).
These ranges are covered with the uncertainty of measurement specified in clause 2. Calibration frequencies lower
than 1 Hz (e.g. 0,4 Hz, which is a reference frequency used in other International Standards) and acceleration
amplitudes smaller than 0,1 m/s2 (e.g. 0,004 m/s2 at 1 Hz) can be achieved using Method 3 specified in this part of
ISO 16063, in conjunction with an appropriate low-frequency vibration generator.
Method 1 (fringe-counting method) is applicable to sensitivity magnitude calibration in the frequency range 1 Hz to
800 Hz and, under special conditions, at higher frequencies (cf. clause 7). Method 2 (minimum-point method) can
be used for sensitivity magnitude calibration in the frequency range 800 Hz to 10 kHz (cf. clause 8). Method 3
(sine-approximation method) can be used for magnitude of sensitivity and phase calibration in the frequency range
1 Hz to 10 kHz (cf. clause 9).
Methods 1 and 3 provide for calibrations at fixed acceleration amplitudes at various frequencies. Method 2 requires
calibrations at fixed displacement amplitudes (acceleration amplitude varies with frequency).