Cardiac defibrillators — Connector assembly DF-1 for implantable defibrillators — Dimensions and test requirements
15. 8. 2002.

Опште информације

90.93     3. 4. 2025.

ISO

ISO/TC 150/SC 6

Međunarodni standard

11.040.40  

engleski   francuski  

Kupovina

Objavljen

Jezik na kome želite da primite dokument.

Apstrakt

This International Standard specifies a unipolar connector assembly, DF-1, intended for use in connecting implantable defibrillator leads to implantable defibrillator generators that do not produce more than 1 kV/50 A peak output. Essential dimensions and performance requirements related to connector fit are specified, along with test methods.
This International Standard does not specify other connector features such as fastening means and material. This International Standard is applicable to the form and fit of the connector assembly, and does not address all aspects of functional compatibility, system performance, or reliability of different implantable defibrillator leads and implantable defibrillator generator assemblies.