It is not always possible to accurately describe the relationship between a set of calibration points with a rectilinear
curve, even by decreasing the working range. Instead of the linear regression analysis, a least-squares fit to a
second-order polynomial is applied (see test for linearity in 4.1.3 of ISO 8466-1:1990[1]). Using this fit, it is possible
to calculate not only the calibration function but also the confidence interval associated with it.
This part of ISO 8466 is intended primarily for use in method development and may not necessarily be applicable to
all routine analyses.
POVUČEN
ISO 8466-2:1993
OBJAVLJEN
ISO 8466-2:2001
90.93
Odluka o potvrđivanju standarda